The MAK 6-2 Plus Sample Gas Conditioner is an essential part of any extractive-gas analyzer system. This unit will continuously dehumidify a sample stream and rapidly separate the condensable liquids from the gas. The MAK 6-2 Plus is ideal for use with SO2, NOx, CO, CO2, or O2 monitors. The cooling system consists of a CFC-free compressor/chiller that operates at over 150% excess capacity. This allows the MAK to compensate for wide variations in input temperature, and guarantees a constant dew point at the outlet. Features include low maintenance, corrosion-resistant Kynar® sample tubing, dual head peristaltic pump for condensate removal, pass line bracket and digital temperature display.
SPECIFICATIONS
Condensate Removal Pump: Dual head peristaltic pump rated @ 26 ml/min
Conditioner Cooling Capacity: 410 BTU/hr (433 kj)
Number of Gas Paths: two
Gas Path Material: PVDF (standard), Stainless Steel (optional), Glass (optional)
Inlet Temperature: PVDF: 284 °F (140 °C), SS & Glass: 356 °F (180 °C)
Gas Path Volume: 8.4 in3 (137 cm3)
Ambient Temperature: 41 °F to 113 °F (5 °C to 45 °C)
Sample Contact Surfaces: PVDF, 316 SS, Teflon
H x W x D: 11.5" x 14.25" x 13.50"
Electrical: 3.18A @ 115V 50/60Hz