Clean Air Engineering Search db1" References About Services Home Equipment
To Speak To A Customer Service Representative Call 800-223-3977
Express Catalog -->MAK 8-2 Conditioner
MAK 8-2
Part Number: 07158

Description

The MAK 8-2 Sample Gas Conditioner is an essential part of any extractive-gas analyzer system. This unit will continuously dehumidify a sample stream and rapidly separate the condensable liquids from the gas. The MAK 8-2 Plus is ideal for use with SO2, NOx, CO, CO2, or O2 monitors. The cooling system consists of a CFC-free compressor/chiller that operates at over 150% excess capacity. This allows the MAK to compensate for wide variations in input temperature, and guarantees a constant dewpoint at the outlet. Features include low maintenance, corrosion-resistant Kynar® sample tubing, dual head peristaltic pump for condensate removal.


Condensate Removal Pump: Dual head peristaltic pump
Max Sample Flow Rate: 2 x 70 l/h
Number of Gas Paths: Two
Gas Path Material: PVDF (standard), Stainless Steel (optional), Glass (optional)
Inlet Temperature: PVDF: 284 °F (140 °C), SS & Glass: 356 °F (180 °C)
Gas Path Volume: 8.4 in3 (137 cm3)
Ambient Temperature: 41 °F to 113 °F (5 °C to 45 °C)
Sample Contact Surfaces: PVDF, 316 SS, Teflon
H x W x D: 14" x 9.5" x 17"
Electrical: 1.3A @ 115V 50/60Hz
Approx. Shipping Weight: 38 lbs.
On Sale Now!
USA List Price Only
To Order Call: 1-800-223-3977
To Speak To A Customer Service Representative Call 800-223-3977
 

Modified: 15-Jun-2006

www.cleanair.com © 2004Clean Air Engineering, All Rights Reserved