| Conditioning
Module Part # 0350
In the condition module, a fraction of the sample passes through
a cooler, where sample moisture content is significantly lowered
by reducing sample temperature.
This process assures that the sample meets the minimum conditions
of the analyzer. The sample is then reheated and mixed with
the original, uncooled sample stream which bypassed the cooler.
The combined stream is brought to 60°C, filtered to remove
condensed particulate, and make available to the GC/MS analyzer
at atmospheric pressure.
Conditioning Module Specifications:
-Sample Flow Rate: 0 to 5 LPM, Excess flows to vent
-Conditioner Type: Solid state Peltier cooler, 316SS construction,
condensate pump
-Conditioning temperature: ambient to .55°C
-Temperature Control: digital PID for temperature indication
-Conditioner flow rate: 0 to 2 LPM
-Secondary particulate filtration: replaceable 0.3 µ
glass filter @ ambient to 204°C
-Flow indication: Precision glass rotometers or -mass flow
meters with digital display
-LxWxH: 15"x11"x15"
-Weight: 43 lbs.
-Power requirements: 110 VAC, 15 amp*
* Probe Module, Conditioning Module and Heated Hose as a system
|
|
Probe
Module Part # 0351
Within the probe module, all components are maintained above
the dew point of the sample typically 120 °C. The sample
is extracted form the stack through a heated probe, the filtered
and pressurized within the probe module. Calibration gas mixtures
are introduced to the probe through a tee and solenoid in
the probe module. The pressurized sample is maintained above
the dew point while traveling through the heated sample line
to the conditioning module. Pressurization insures that any
system leaks will not result in sample dilution or loss.
Probe Module Specifications:
-Sample Temperature: 204 °C to 1200°C (with High temp
probe)
-Sample Moisture Content: 2 to 50% by volume
-Gas Pressure: +/- 16"
-Sample Flow Rate: 0 to 5 LPM
- Sample Exit Pressure: 0 to 15 psig
-Pressure Pump Construction: heated 316SS head diaphragm pump
-Probe Construction: replaceable heated 1/4" liner, 1"
OD outer sheath, 316 SS construction
-Particulate filtration: replaceable 0.3µ glass fiber
filter
-Operating temperature: ambient to 204°C
-Temperature Control: Digital PID
-Weight: 24 lbs. |