Clean Air Engineering Search db1" References About Services Home Equipment
To Speak To A Customer Service Representative Call 800-223-3977
Express Catalog -->Sample Delivery Conditioning System
SDCS
Part Number:0350&0351



Expand the range of applications for your portable GC/MS and INFICON® HAPSITE® with a sample delivery and conditioning system that handles hostile gas environments and gives you reliable results the first time,
on-site, immediately. This versatile system - the Clean Air Engineering Sample Delivery and Conditioning System (SDCS) - allows the environmental engineers to measure hot, wet samples. No other portable GC/MS has an interface system which can sample directly into a smoke stack, and no other system offers real time analysis. Why send samples
to a lab, then wait two to three weeks for results, when the SDCS provides results in two to three minutes? This low bias instrument conditions the undiluted sample, delivering a gas the analyzer can handle, without changing the content or characteristics of the sample.

Features
Allows for instantaneous results
Eliminates additional laboratory costs
Increases efficiency with a portable, easy to handle system
Eliminates requirements for complete moisture removal
Minimizes contact between the sample gas and condensate
Eliminates leaks and sample dilution via sample pressurization
Provides total control over sample conditioning process
Enables you to quantify gas concentrations to establish sample rates for VOST testing

 

Conditioning Module Part # 0350

In the condition module, a fraction of the sample passes through a cooler, where sample moisture content is significantly lowered by reducing sample temperature.
This process assures that the sample meets the minimum conditions of the analyzer. The sample is then reheated and mixed with the original, uncooled sample stream which bypassed the cooler. The combined stream is brought to 60°C, filtered to remove condensed particulate, and make available to the GC/MS analyzer at atmospheric pressure.

Conditioning Module Specifications:

-Sample Flow Rate: 0 to 5 LPM, Excess flows to vent
-Conditioner Type: Solid state Peltier cooler, 316SS construction, condensate pump
-Conditioning temperature: ambient to .55°C
-Temperature Control: digital PID for temperature indication
-Conditioner flow rate: 0 to 2 LPM
-Secondary particulate filtration: replaceable 0.3 µ glass filter @ ambient to 204°C
-Flow indication: Precision glass rotometers or -mass flow meters with digital display
-LxWxH: 15"x11"x15"
-Weight: 43 lbs.
-Power requirements: 110 VAC, 15 amp*
* Probe Module, Conditioning Module and Heated Hose as a system

 

 

Probe Module Part # 0351

Within the probe module, all components are maintained above the dew point of the sample typically 120 °C. The sample is extracted form the stack through a heated probe, the filtered and pressurized within the probe module. Calibration gas mixtures are introduced to the probe through a tee and solenoid in the probe module. The pressurized sample is maintained above the dew point while traveling through the heated sample line to the conditioning module. Pressurization insures that any system leaks will not result in sample dilution or loss.

Probe Module Specifications:

-Sample Temperature: 204 °C to 1200°C (with High temp probe)
-Sample Moisture Content: 2 to 50% by volume
-Gas Pressure: +/- 16"
-Sample Flow Rate: 0 to 5 LPM
- Sample Exit Pressure: 0 to 15 psig
-Pressure Pump Construction: heated 316SS head diaphragm pump
-Probe Construction: replaceable heated 1/4" liner, 1" OD outer sheath, 316 SS construction
-Particulate filtration: replaceable 0.3µ glass fiber filter
-Operating temperature: ambient to 204°C
-Temperature Control: Digital PID
-Weight: 24 lbs.

Price: $17,000
USA List Price Only
To Order Call: 1-800-223-3977
To Speak To A Customer Service Representative Call 800-223-3977
 

Modified: 15-Jun-2006

www.cleanair.com © 2004Clean Air Engineering, All Rights Reserved
Web site and legal information