|
Measurement
of Gaseous Hydrogen Chloride Emissions at Portland Cement Kilns
by Fourier Transform Infrared (FTIR) Spectroscopy – USEPA
Method 321
USEPA Method 321 is designed for the application of Fourier Transform
Infrared Spectroscopy (FTIR) in extractive measurements of gaseous
HCl concentrations in portland cement kiln emissions.
This method applies to the measurement of hydrogen chloride (HCl)
concentrations both before and after particulate matter control
devices. It also applies to either continuous flow through
measurement (with isolated sample analysis) or grab sampling (batch
analysis). HCl is measured using the mid-infrared spectral
region for analysis (about 400 to 4000 cm-1 or 25 to 2.5 µ m).
The analytical range is determined by the instrumental design and
the composition of the gas stream. For practical purposes
there is no upper limit to the range because the pathlength may
be reduced or the sample may be diluted. The lower detection
range depends on:
- the absorption
coefficient of the compound in the analytical frequency region;
- the spectral
resolution;
- the interferometer
sampling time;
- the detector
sensitivity and response;
- the absorption
pathlength.
The practical
lower quantification range is usually higher than the instrument
sensitivity allows and is dependent upon:
- the presence
of interfering species in the exhaust gas including water (H2O),
carbon dioxide (CO2), and sulfur dioxide (SO2);
- analyte
losses in the sampling system;
- the optical
alignment of the gas cell and transfer optics;
- the quality
of the reflective surfaces in the cell (cell throughput).
Under typical
test conditions (moisture content of up to 30% and CO2
concentrations from 1 to 15 percent), a 22 meter path length cell
with a suitable sampling system achieves a lower quantification
range of from 1 to 5 ppm for HCl.
Sample transport presents the principal difficulty in directly measuring
HCl emissions. This identical problem is overcome by any extractive
measurement method. HCl is reactive and water soluble.
The sampling system is adequately designed to prevent sample condensation
in the system.
Potential interferences include:
- analytical
interferences
- background
intereferences
- spectral
interferences
- sampling
system interferences.
This method
will not measure HCl under conditions:
- where the
sample gas stream condenses in the sampling system or the instrumentation;
- where a
high moisture content sample relative to the analyte concentrations
imparts spectral interference due to the water vapor absorbance
bands.
|